| Sign In | Join Free | My gastesters.com |
|
Brand Name : JUGE
Place of Origin : Nanjing, China
Model Number : AG9258C
Certification : IOS9001, CE
MOQ : 1 Unit
Price : USD2000-7000
Packaging Details : Neutral carton
Delivery Time : 5-8 working days
Payment Terms : Online order,T/T
Supply Ability : 100 Unit/Week
Measurement Principle : DPD colorimetric method
Range : 0-5 mg/L (ppm)
Resolution : 0.001 mg/L (ppm)
Accuracy : ±5%
Cycle Time : Adjustable from 5-9999 min (factory default: 5 min)
Display : 7-inch color LCD touchscreen
Dimensions : 350×450×200mm
Weight : 11.0kg
Online Colorimetry DPD Wall-mounted Residual Chlorine FCL Analyzer
>>Overview
This instrument utilizes the DPD colorimetric method to detect residual chlorine concentration. It offers precise measurement, cost-effectiveness, and low maintenance, making it suitable for residual chlorine monitoring in chlorination disinfection processes and drinking water distribution networks. With a 30-day continuous reagent supply enabling unattended operation, it significantly reduces operational and maintenance costs. Ideal for residual chlorine measurement in tap water, circulating cooling water, wastewater treatment chlorination processes, and municipal water pipeline networks.
>>Features
▪ Wide power input, 7-inch touchscreen design.
▪ DPD colorimetric method for more accurate and stable measurements.
▪ Adjustable measurement cycle.
▪ Automatic measurement and cleaning.
▪ External signal input to control measurement start/stop.
▪ Selectable automatic or manual mode.
▪ 4-20mA and RS485 outputs, relay control.
▪ Data storage with USB export support.
>>Technical Parameter
| Measurement Principle |
DPD colorimetric method |
| Measurement Range |
0-5 mg/L (ppm) |
| Resolution |
0.001 mg/L (ppm) |
| Accuracy |
±5% |
| Cycle Time |
Adjustable from 5-9999 min (factory default: 5 min) |
| Display |
7-inch color LCD touchscreen |
| Power Supply |
110-240VAC, 50/60Hz; or 24VDC, 20W |
| Analog Output |
4-20mA output, Max. 750Ω |
| Digital Communication |
RS485 Modbus RTU |
| Alarm Output | 2 sets of contacts: (1) Sampling signal control (2) Hi/Lo alarm points with hysteresis setting, 5A/250VAC, 5A/30VDC |
| Data Storage |
Historical trends and 2-year data storage, supports USB export |
| Operating Conditions |
Temperature: 0-50°C; Humidity: 10-95%, non-condensing |
| Protection Rating |
IP65 |
| Dimensions |
350×450×200mm |
| Weight |
11.0kg |
| Flow Requirements |
Recommended flow rate: 300-500 mL/min; Pressure: 1 bar |
|
|
Online Colorimetry DPD FCL Residual Chlorine Analyzer Wall Mounted Images |